当前位置:网站首页 > 论文专利 > 论文
【论文专利】

题目/Title:A Neuro-Stimulus Chip with Telemetry Unit for Cochlear Implant

作者/Author:
                        Mian Dong,Chun Zhang,Zhihua Wang,Dongmei Li

会议/Conference:BioCAS 2004

地点/Location:Singapore

年份/Issue Date:2004.1-3 Dec.

页码/pages:pp. S1/3/INV - S1/39-12

摘要/Abstract:
This paper presents a 16-channel cochlear implant which supports 4 stimulus modes and provides a highest stimulus rate of 15k pulses per second per channel, and dynamic range of 60dB with the 10 bit D/A converter. The device consists of a coil, a rectifier and regulator, and a mixed analog and digital neuro-stimulus ASIC which decodes telemetry command and generates stimulus signals. The chip is fabricated using 1.2μm 2M/2P standard CMOS process and the die size is 3.56 mm × 3.67mm. The circuit is working properly at a telemetry data rate of 500 Kbps in carrier of 10 MHz while the peak power consumption is 35 mW.

全文/Full text:PDF